User information / operating instructions for CAMOped

 instruction manual CAMOped instruction manual
Device assembly

Procedure
Insert the injured leg
Insert the healthy leg
Start the exercise

Operation
Pre-adjustment of the CAMOped device
Positioning of the CAMOped

Operating elements and basic settings
Angle limiters
Unlocking the sliding device
Flexion/extension support
Varus/valgus locking mechanism
Ankle joint extension resistance
Adjustment towards length of lower leg
Resistance adjustment
Telescopic reinforcement
Removing the shell
Pre-adjusting movement resistance (not to be done by patient - authorized personnel only)

 

Device assembly

1 Angle limiter     9 Ankle joint resistance adjustment
2 Pads     10 Closure strap
3 Slider release mechanism     11 Flexion-extension support locking mechanism
4 Leg shell loosening mechanism     12 Varus/Valgus locking mechanism
5 Side screen closure     13 Adjusting screw for telescopic reinforcement
6 Resistance adjustment     14 Calf shell
7 End rubber     15 Side screen
8 Telescopic reinforcement     16 Foldable supports

 
 Video angle limiters Video angle limiters Video angle limiters Video angle limiters

Angle limiters

    Use to restrict the range of motion regarding flexion and extension of the legs. To adjust the range of movement, press the stops outwards and set the limiters as recommended by the physician.
Secure the angle by clicking the stops into position. (Adjust both sides symmetrically).

Basic setting:
Upper stop 9 (approx. 90°)
Note: The numbers on the angle limiters are only hinting towards the position of the knee. Accurate angles must be measured on the patient's knee.

 
 Video sliding device Video sliding device Video sliding device Video sliding device

Unlocking the sliding device

    When activating the slider release mechanism, sliders can be adjusted to the length of the patient's legs.

Basic setting:
Press the "banana-shaped button" and move the sliders into the desired postion. Release the "banana-shaped button"

 

Flexion/extension support

    Use to address either flexion or extension deficits. The further the leg shell is positioned in the flexion direction, the greater the support for flexing the leg, and thus the greater the activation of the quadriceps muscle for extension.
The further the splint is pushed in the extension direction, the greater the extension support.
(Adjust both sides symmetrically)

Basic setting:
Position 0

 

Varus/valgus locking mechanism

    Allows for locking of the leg shells in varus/valgus direction. Locking is recommended only in the case of patients with a poor physical constitution.

Basic setting:
Free natural movement of the knee to promote proprioception.
Rotating button not used - the shell remains mobile.

 
 Video ankle joint extension resistance Video ankle joint extension resistance Video ankle joint extension resistance Video ankle joint extension resistance

Ankle joint extension resistance

    Changes the resistance against extension in the ankle joint.
Resistance can be altered by varying the tension on the orange rubber band.
Rubber band in lowest catcher - least resistance
Rubber band in highest catcher - most resistance

Basic setting:
Rubber band inserted in the lowest catcher (least resistance)

 

Adjustment towards length of lower leg

    The calf shell can be moved lengthways in order to ensure optimal positioning of the lower leg.
Hold the shell firmly on the foot section and slide calf support into desired position.

 
 Video resistance Video resistance Video resistance Video resistance

Resistance adjustment

    The device can be set to the desired movement resistance by turning the positioning wheel.
The set value (0-3) can be read in the position indicated.

Basic setting:
0 (no resistance)

Position 1 2 3
Resistance ca. (N) 10 15 30

 

Telescopic reinforcement

    In order to ensure optimal stability, the device can be set against the end of the bed or a wall using the telescopic reinforcement.

 

Removing the shell

    To remove the lower leg shell, press the shell unlocking mechanism and take off the shell.
When inserting the shell, it must audibly click into place.

 

Pre-adjustment of the CAMOped device

  • Open the straps on the leg shells.
  • Adjust the angle limiters to accommodate the range of motion for extension/flexion, as directed by the doctor. If no instructions are given, keep basic adjustment (0/ 9). The range of movement is read off between the stops. The same adjustment must be made on either side.
  • LCenter the sliding device (see basic positioning on P.6).
  • Adjust the extension/flexion support as directed by the doctor. Set to 2 if no instructions are given.
  • Adjust the varus/valgus locking mechanism as instructed by the doctor. Do not lock if no instructions are given.
  • Adjust the resistance to movement as directed by the doctor. If no instructions are given, do not use resistance (setting 0).

 Video for Application Video for Application Video for Application Video for Application

Positioning of the CAMOped

    The patient should be level with the CAMOped device and preferably in the seated position (e.g. on the floor with back against the couch).
  • Place the CAMOped between the patient's legs
  • Distance between device and patient's crotch 1 to 2 inches
  • Unfold the CAMOped legs to stabilize the device
If required:
To increase the stability of the device, use the telescopic reinforcement and set the device against e.g. a wall.

 
 Video for Application Video for Application Video for Application Video for Application

Insert the injured leg

    Insert the injured leg in a rather stretched position.
Note!

Check beforehand that the angle limiters are adjusted according to the physician's recommendations and clicked into position.
Otherwise the injured leg can be overstretched.

 
 Video for Application Video for Application Video for Application Video for Application

Insert the healthy leg

    The injured leg remains stretched whilst the healthy leg is bent. Close the straps.

If necessary, adjust to user's leg length by positioning the sliding device: Press the slider release mechanism ("banana-shaped button") and lift and bend the healthy leg according to the physician's recommendation.

 
 Video for Application Video for Application Video for Application Video for Application

Start the exercise

    Option: Resistance can be adjusted (see "Resistance adjustment")

 

Pre-adjusting movement resistance (not to be done by patient - authorized personnel only)

    The resistance settings can be pre-adjusted using a nut (13mm). The device is correctly adjusted when the sliding mechanism can move unhindered in position 0 and gentle resistance is perceptible in position 1.

The brake disc should be changed by a specialist.

Maintenance recommendation: The apparatus should be thoroughly overhauled at least once a year. Your retailer will be pleased go give you information about the service options available (also see service information). .